Thermodynamics Research Center / ThermoML | Thermochimica Acta

Temperature controlled photoacoustic device for thermal diffusivity measurements of liquids and nanofluids

Agresti, F., Ferrario, A., Boldrini, S., Miozzo, A., Montagner, F., Barison, S., Pagura, C., Fabrizio, M.
Thermochim. Acta 2015, 619, 48-52
ABSTRACT
In this paper, the design and development of a device for thermal diffusivity measurement of fluids exploiting the photoacoustic effect is presented. Compared to similar devices working only at room temperature, the version of the apparatus we present allows precise measurements of thermal diffusivity of fluids from room temperature to 70.C. The reliability of the measurements is shown by the estimation of thermal diffusivity of well-known test liquids. The possibility to detect slight thermal diffusivity variations (around 1%) is verified by measurements on alumina based water nanofluids showing remarkable advantages in respect to alternative techniques.
Compounds
# Formula Name
1 H2O water
2 C2H6O2 1,2-ethanediol
Datasets
The table above is generated from the ThermoML associated json file (link above). POMD and RXND refer to PureOrMixture and Reaction Datasets. The compound numbers are included in properties, variables, and phases, if specificied; the numbers refer to the table of compounds on the left.
Type Compound-# Property Variable Constraint Phase Method #Points
  • POMD
  • 1
  • Thermal diffusivity, m2/s ; Liquid
  • Temperature, K; Liquid
  • Pressure, kPa; Liquid
  • Liquid
  • Photoacoustic measurements
  • 5
  • POMD
  • 2
  • Thermal diffusivity, m2/s ; Liquid
  • Temperature, K; Liquid
  • Pressure, kPa; Liquid
  • Liquid
  • Photoacoustic measurements
  • 5